CT Quality Control

download CT Quality Control

of 35

Transcript of CT Quality Control

  • 8/3/2019 CT Quality Control

    1/35

    QC for CT-Simulator

    In this tour we will check the following :1. Image quality

    2. CT Dose

    3. Mechanical performance

  • 8/3/2019 CT Quality Control

    2/35

    Image quality

  • 8/3/2019 CT Quality Control

    3/35

    Head Foot

    4 cm

    20cm

    ModelModelModelModel

    Image quality

  • 8/3/2019 CT Quality Control

    4/35

    Model 1alignment , slice thickness and CT

    numberMust see all fourBBs(in Modules 1 & 4)

    Longer wire musthave same numberof lines above andbelow (1)

    Wires are 0.5 mmapart in z-direction

    Image quality

  • 8/3/2019 CT Quality Control

    5/35

    CT Number Calibration

    According to NCRP No.99

    Water CT# = 0 2

    Air CT# = -1000 3

    Image quality

    Polyethylene -97 HUWater 0 HUAcrylic +120 HUBone +910 HU

    Air -1000 HU

  • 8/3/2019 CT Quality Control

    6/35

    Image quality

  • 8/3/2019 CT Quality Control

    7/35

    Image quality

    Model 2

    The low contrast phantomThere are four cylinders foreach of the following

    diameters:2 mm, 3 mm, 4mm, 5 mmand 6mm.The space between eachcylinder is equal to the

    diameter of the cylinder.

  • 8/3/2019 CT Quality Control

    8/35

    Image quality

    View the best image

    A: Record the mean counts in the ROI inside the 25mm rod

    B: Record the mean counts in the ROI outside the 25 mm rodSD: the Standard Deviation from the ROI outside the 25 mm rod

    CNR = |A-B|/SD 1

  • 8/3/2019 CT Quality Control

    9/35

    Image quality

    The noise can be assessed by scan the phantom andcalculate the standard deviation CT of the HUs in aregion-of-interest (ROI)

    Noise is expressed relative to the linear attenuation

    coefficient of water

    wm

    wm

    CTCTCS

    w

    ROICSSD

    Noise

    %100

    Mean = 0 3

    SD = 3.4 0.4

    Based on GE operator manual

    Noise

  • 8/3/2019 CT Quality Control

    10/35

    Module 3

    uniformity and distance

    Image quality

  • 8/3/2019 CT Quality Control

    11/35

    Image quality

    The uniformity can be assessed by measuring the average HU in

    smaller ROIs about 5 mm in radius that are located in the center andperiphery of the phantom ( 3 , 6 , 9, 12 oclock positions ). The largest

    difference between any peripheral HU and the central HU is

    determined.

    Image uniformity

    CT number variation for Water equal 0 5 HU

  • 8/3/2019 CT Quality Control

    12/35

    two small targets fortesting in-plane

    distancemeasurementaccuracy.

    Image quality

  • 8/3/2019 CT Quality Control

    13/35

    Image quality

    This phantom is used toassess high contrast (spatial)resolution. It contains eightbar resolution patterns: 4, 5,6, 7, 8, 9, 10 and 12 lp/cm .

    Model 4

    High contrast resolution

  • 8/3/2019 CT Quality Control

    14/35

    12

    10

    9

    8

    7

    6

    5

    4

    Image quality

  • 8/3/2019 CT Quality Control

    15/35

    Image quality

    artifacts

  • 8/3/2019 CT Quality Control

    16/35

    2 standard CT dosimetry phantoms were adopted by theFood and Drug Administration and still are used today.contain several 1-cm-diameter holes for insertion ofdosimeters.

    CTDI Phantoms32cm diam. X 15cm16cm diam. x 15cm

    CT - Dose

    CT Dose

  • 8/3/2019 CT Quality Control

    17/35

    edge

    edge

    Middle

    Middle

    100mm Ion Chamber

    CTDI Phantoms32cm diam. X 15cm16cm diam. x 15cm

  • 8/3/2019 CT Quality Control

    18/35

    Kel : electrometer reading calibration factor

    Ctp : temperature pressure correction factor

    Nx :chamber exposure correction factor (R/c)

    f: exposure-to-dose conversion factor, D= fX

    100 mm :ionization chamber length

    beam width = n T

    )(..

    )(100.100

    mmwidthbeamtotal

    mmfNCKreadingmeterCTDI

    xtpel

    Procedure simply is :

    Phantom is place in the center ofimaging plane and aligned withcentral axis of the scanner bylasers and single scan isacquired .

    CT Dose

  • 8/3/2019 CT Quality Control

    19/35

    Tolerance

    Reference

    SourceTolerance limitsTest objective

    Performance

    Parameter

    AAPM Report

    Task Group 66

    20% of

    manufacturerspecifications

    verify safe dose

    delivered fromthe scanner

    Patient dose from

    CT-scan

    CT Dose

  • 8/3/2019 CT Quality Control

    20/35

    The CTDI varies across the field of view (FOV). For example,for body CT imaging, the CTDI is typically a factor or twohigher at the surface than at the center of the FOV.

    Weighted CDTIW

    The average CTDI across the FOV is estimated by theWeighted CTDI (CTDIw)

    CT Dose

  • 8/3/2019 CT Quality Control

    21/35

    CT Dose

  • 8/3/2019 CT Quality Control

    22/35

    pitch is defined as the ratio of the table travel perrotation (I) to the total beam width (Nx T)

    it does not indicate the total energy deposited into the scan

    volume because it is independent of the length of the scan. Thatis, its value remains unchanged whether the scan coverage is

    10 or 100 cm. It estimates the dose for a 100-mm scan length

    only, even though the actual volume-averaged dose will

    increase with scan length

    CTDIvol is a useful to compare CT protocols betweenscanners

    CT Dose

  • 8/3/2019 CT Quality Control

    23/35

    The DLP reflects the total energy absorbed attributableto the complete scan acquisition , and thus thepotential biological effect .

    Dose-Length Product (DLP)

    DLP = CTDIvol * L , where L is the Z-length of the scan

    CT Dose

  • 8/3/2019 CT Quality Control

    24/35

    Limitation

    For body scan lengths of 250 mm or more, theaccumulated dose closely approaches the limiting

    equilibrium dose. However, CTDI100 underestimatesthe equilibrium dose CTDI (or MSAD for pitch of unity)by a factor of ~ 0.6 on the central axis and by ~ 0.8 onthe periphery .therefore the DLP by factor of ~ 0.7 .

    In order to measure the equilibrium dose, a bodyphantom length of ~ 400 mm is required.

    CT Dose

  • 8/3/2019 CT Quality Control

    25/35

    Mechanical performance

    Positioning lasersWe can perform this test in two ways

    By using phantom and BBs on up , down , left and right

    Mechanical

  • 8/3/2019 CT Quality Control

    26/35

    Using film

    Mechanical

  • 8/3/2019 CT Quality Control

    27/35

    Using phantom

    By making pin holes mark along illuminateLine .The deviation should be within 2 mm

    Mechanical

  • 8/3/2019 CT Quality Control

    28/35

    Mechanical

    On the displayed CT image,the four (1mm) holes should be visible.

    On Film,

    we measure the deviation of pin-hole center from scanplane center. The deviation should be within 2 mm

    Reference Source:

    AAPM Report No. 39 and GE Operator Manual

    Tolerance

  • 8/3/2019 CT Quality Control

    29/35

    Mechanical

    Couch deflection

  • 8/3/2019 CT Quality Control

    30/35

    Mechanical

    Couch deflection

  • 8/3/2019 CT Quality Control

    31/35

    Mechanical

    Couch deflection

  • 8/3/2019 CT Quality Control

    32/35

    Mechanical

    Couch height

  • 8/3/2019 CT Quality Control

    33/35

    Mechanical

    Couch height

  • 8/3/2019 CT Quality Control

    34/35

    Mechanical

    Couch distance accuracy

    C

  • 8/3/2019 CT Quality Control

    35/35

    Mechanical

    Couch distance accuracy