Four Channels Data Acquisition System for Silicon Photomultipliers
Mateusz Baszczyk, Piotr Dorosz, Sebastian Głąb,Wojciech Kucewicz, Łukasz Mik, Maria Sapor
Department of Electronics,AGH – University of Science and Technology,Al. Mickiewicza 30, 30-059 Krakow, Polande-mail kucewicz@agh.edu.plsglab@agh.edu.pl
Agenda
• Fluorescence;• Silicon Photomultiplier description;• Description of previous measurement system;• Four channel data acquisition system;• Results;• Summary;• Future work.
2
Why do we measure fluorescence light intensity?
• Dye-labelled cells can be count by fluorescence light intensity measurement;
• At low concentrations the fluorescence intensity is generally proportional to the concentrations of fluorophore.
3
Fluorescein and Bio-Tek filters spectra
fluorescein resorufin
4
Optical block
Blue LED
S500LLB4G-H+
485/20nm optical filter
1.5 ml cuvette with fluorescein solution
530/25nm optical filter
5mm diameter optical channel
5mm diameter optical collimator
and fiber cable with 940um core
5
Silicon Photomultiplier
The idea appears in the late 90-ties:•„Photon Detection with High Gain Avalanche Photodiode Arrays” S. Vasile, P. Gothoskar, R. Farrell, D. Sdrulla•„Silicon avalanche photodiodes on the base of metal-resistor-semiconductor (MRS) structures” V. Saveliev, V. Golovin 6
Structure
7
Current - Voltage Characteristic
8
Photodiode Operation Mode
1 photon generates 1 electron-hole pair9
Avalanche Operation Mode
1 photon generates 10-200 electron-hole pairs10
Geiger Operation Mode
1 photon generates 105 - 106 electron-hole pairs11
Typical SiPM Parameters
sensL s1020 Hamamatsu S10362-11-025C
Spectral Range (λ) 400-1100 nm 320-900 nm
Peak Spectral Response (λ) 490 nm 440 nm
Breakdown Voltage (Vbr) 28.6 V ---
Operating Voltage (V) 30.6 V 70.82 V
Microcell Gain 106 2.75 · 105
SPM Pixel Active Area Φ 1 mm Φ 1 mm
Number of Pixels 400 1600
Dark Rate (MHz) 0.42 0.3
12
Previous Measurement System For Fluorescence
Picosecond 2600C
generator
Cuvette holderwith
specimen
GPAamplifier
QDC device
PC with Labview
application
excitationfilter
emissionfilter
Agilent33250
generator
trigger signal gate pulses
13
First Measurement System For Fluorescence
14
Acquisition System
15
Acquisition System
16
Front-end ASIC
17
Front-end ASIC
18
Front-end ASIC
19
Linearity – high gain
20
Linearity – low gain
21
Measurement results
22
Measurement results
23
Measurement results
......),,,(22
22
21
21
22
21111
mxmx
eaeamaxf 24
Measurement results
25
Measurement results
26
Measurement results
27
Light Intensity
28
Light Intensity
29
Light Intensity
30
Without Coincidence
31
Coincidence
32
Results of fluorescence light measurements
976.6 [pg/ml]61.05 [pg/ml]
y = -572500 x² + 223.6 x + 6458000 – fluoresceiny = -373800 x² + 170.3 x + 14030000 – resorufin
33
Influence of Temperature and Bias
34
Dark Current Rate
35
Influence of Temperature and Bias
36
Conclusions
• We have designed, built and optimized readout system for detection of single photons;
• We have tested readout circuit with light and confirmed that it is linear;
• We can distinguish a dozen or so photons and eliminate thermal generation from measurement data;
• System with silicon photomultiplier has a good sensitivity and can be used to measure fluorescence light intensity for fluorescein concentration from 1ug/ml to about 60pg/ml and resorufin at the level of 1ng/ml up to 1ug/ml.
37
Future work
• Stabilization system compensating temperature influence on SiPM gain by varying supply voltage;
• Measurements of fluorescent dyes in microbioreactors.