SSI-008

download SSI-008

of 4

Transcript of SSI-008

  • 7/28/2019 SSI-008

    1/4

    Summary

    Sensitivity was tested for four steroids on the LCMS-

    8030 triple quadrupole mass spectrometer coupled

    with a Nexera UHPLC.

    Background

    Steroids are powerful hormones that are needed fornormal biological activity but may also be present in

    the environment. Water contaminated with steroids

    poses a serious environmental and public health

    threat. Sensitive and rapid methods are needed for

    accurate quantitative analysis. UHPLC-MS-MS meth-

    ods for analysis of these compounds were developed

    and the sensitivity for each analyte was tested.

    Method

    Standards for four steroids were obtained at a concen

    tration of 1 mg/mL in methanol. The

    standards were: Testosterone (T4), Progesterone (P4

    Estradiol (E2), and Ethinylestradiol (EE2). Standards

    were diluted in 50/50 water/acetonitrile for optimizatio

    of mass spectrometry parameters.

    Multiple Reaction Monitoring (MRM) was used for

    analysis. Collision energy, quadrupole prerod bias

    voltage, and other MS parameters were optimized by

    flow injection of the individual standards. T4 and P4

    ionized most efficiently using a 0.1% formic

    acid/acetonitrile mobile phase using the DUIS dual io

    Figure 1: Structures of tested steroidsTable 1: Dilution levels (ng/mL)

    Table 2: MRM parameters

    Testosterone Progesterone

    Estradiol Ethinylestradiol

    O

    OH

    H

    H

    H

    O

    O

    H

    H

    H

    HO

    OH

    H

    H

    H

    HO

    OH

    H

    H

    H

    Type Event# +/- Compound Name (m/z) Dwell Time (msec) Q1 Pre Bias(V) CE Q3 Pre Bias(V)

    MRM 1 + Testosterone 288.9>97 80 -14 -25 -10

    MRM 2 + Progesterone 315.4>109.1 80 -12 -30 -20

    Type Event# +/- Compound Name (m/z) Dwell Time (msec) Q1 Pre Bias(V) CE Q3 Pre Bias(V)

    MRM 1 - Estradiol 271.4>183.1 50 13 41 30

    MRM 2 - Ethinylestradiol 295.4>145 50 14 47 28

    Measurement Time (min)

    0-2

    0-2

    Measurement Time (min)

    0-2

    0-2

    Level T4 P4 E2 EE2

    L1 370.3 370.3 10000 10000

    L2 123.5 123.5 3333 3333

    L3 41.17 41.17 1111 1111

    L4 13.72 13.72 370.3 370.3

    L5 4.57 4.57 123.5 123.5

    L6 1.52 1.52 41.17 41.17

    L7 0.582 0.582 13.72 13.72

    L8 4.57 4.57

    Steroids: Testosterone, Progesterone, Estradiol,

    Ethinylestradiol

    ApplicationNews

    No. SSI-008

    Liquid Chromatography Mass Spectrometry

  • 7/28/2019 SSI-008

    2/4

    spray (ESI with corona discharge) source in positive

    mode. The MRM transition ofm/z288.997 was

    monitored for T4 and the transition ofm/z315.4109.1 was used for P4.

    E2 and EE2 ionized most efficiently using a

    water/acetonitrile mobile phase with atmospheric

    pressure chemical ionization (APCI) in negative mode.

    Therefore these compounds were analyzed in a

    separate method. The MRM transitions for E2 and EE2

    were 271.4183.1 and 295.4145.0 respectively.

    A Shimadzu Shimpack-XR ODS III column (1.6 m, 2

    50 mm) was used for analysis. A binary gradient of

    0.1% formic acid and acetonitrile was used for T4 an

    P4 while a binary gradient of water and acetonitrilewas used for E2 and EE2. The gradient began at 55

    acetonitrile and increased to 95% over 1 minute. Aft

    a 6 second isocratic hold, the column was

    re-equilibrated for approximately 1 minute. The colu

    temperature was 40 C and the injection volume wa

    10 L.

    Calibration curves were prepared by serial dilution o

    mixed standard as shown in Table 1.

    110504.SetEPAx01_

    15_

    EPAMIX2L10_

    015.lcd

    P4 r2 = 0.999 T4 r 2 = 0.991

    Figure 2: Calibration curves and selected mass chromatograms for P4 and T4

    P4 Level 7

    P4 Level 2

    T4 0.065 ng/mL

    T4 Level 2

    0 100 200 300 Conc.0.0

    1.0

    2.0

    3.0

    Area(x100,000)

    0 100 200 300 Conc0.00

    0.25

    0.50

    0.75

    1.00

    Area(x1,000,000)

    0.0 0.5 1.0 1.5

    0.0

    1.0

    2.0

    3.0

    4.0

    5.0(x100)

    1:315.40>109.10(+)

    0.

    992

    0.5 1.0 1.5

    0.0

    0.5

    1.0

    1.5

    (x100)2:288.90>97.00(+)

    0.

    616

    110504.SetEPA

    x01_

    8_

    EPAMIX2L12_

    008.lcd

    0.0 0.5 1.0 1.5

    0.00

    0.25

    0.50

    0.75

    1.00

    (x100,000)1:315.40>109.10(+)

    0.

    998

    110504.SetEPAx01_

    29_

    EPAMIX2L05_

    029.lcd

    110504.SetE

    PAx01_

    29_

    EPAMIX2L05_

    029.lcd

    0.5 1.0 1.5

    0.0

    0.5

    1.0

    1.5

    2.0

    2.5

    3.0(x100,000)

    2:288.90>97.00(+)

    0.

    620

    110504.SetEPAx01.

    lcb

    110504.SetEPAx0

    1.lcb

    pplicationews

    No.SSI-008

  • 7/28/2019 SSI-008

    3/4

    Results and Discussion

    Representative chromatograms of the T4/P4 andE2/EE2 methods are shown in Figures 2 and 3. The

    limit of detection (LOD, S/N > 3) for P4 was between

    0.194 and 0.582 ng/mL using a 10 L injection, and the

    limit of quantitation (LOQ, S/N >10) was approximately

    0.5 ng/mL. For T4, the LOD was approximately

    0.065 ng/mL and the LOQ was approximately

    0.2 ng/mL. For T4 and P4, the calibration curve was

    constructed between 0.194 and 370 ng/mL. The curves

    were linear in the tested range (r2 > 0.99).

    For E2, the LOD was approximately 1.5 ng/mL and th

    LOQ was approximately 4.5 ng/mL. The LOD and LOfor EE2 was similar. The calibration curves were

    constructed from 4.57 ng/mL up to 10 g/mL, and the

    curves were both linear (r2 > 0.99)

    Conclusion

    Rapid and sensitive methods for

    determination of four steroids was developed using

    UHPLC-MS-MS.

    E2 r2 > 0.999 EE2 r2 > 0.999

    Figure 3: Calibration curves and selected mass chromatograms for E2 and EE2

    E2 1.52 ng/mL

    E2 Level 3

    EE2 1.52 ng/mL

    EE2 Level 3

    0 2500 5000 7500 Conc.0.0

    1.0

    2.0

    Area(x100,000)

    0 2500 5000 7500 Conc.0.0

    0.5

    1.0

    1.5

    Area(x100,000)

    110505.SetEP

    Ax02.lcb

    110505.SetEPA

    x02.lcb

    0.0 0.5 1.0 1.5

    0.0

    1.0

    2.0

    3.0

    4.0

    (x10)1:271.40>183.10(-)

    0.

    209

    0.0 0.5 1.0 1.5

    0.0

    0.5

    1.0

    1.5

    2.0

    2.5

    (x10,000)1:271.40>183.10(-)

    0.

    214

    110

    505.SetEPAx02_

    30_

    E2EE2L03_

    030.lcd

    110505.SetEPAx02_

    30_

    E2EE2L03_

    030.lcd

    110505.S

    etEPAx02_

    11_

    E2EE2L09_

    011.lcd

    0.5 1.0 1.5

    0.00

    0.25

    0.50

    0.75

    1.00

    1.25

    1.50(x10,000)

    2:295.40>145.00(-)

    0.

    234

    0.5 1.0 1.5

    0.0

    0.5

    1.0

    1.5

    2.0

    2.5

    (x10)2:295.40>145.00(-)

    0.

    234

    110505.SetE

    PAx02_

    10_

    E2EE2L09_

    010.lcd

    pplicationews

    No.SSI-008

  • 7/28/2019 SSI-008

    4/4

    SHIMADZU SCIENTIFIC INSTRUMENTS

    7102 Riverwood Drive, Columbia, MD 21046, USA

    Phone: 800-477-1227/410-381-1227, Fax: 410-381-1222

    URL: www.ssi.shimadzu.com

    For Research Use Only. Not for use in diagnostic procedures.

    The contents of this publication are provided to you as is without warranty of any kind, and are subject to change without notice. Shimadzu does

    not assume any responsibility or liability for any damage, whether direct or indirect, relating to the use of this publication.

    Shimadzu Corporation, 201

    First Edition: November 20

    www.shimadzu.com/an/

    pplicationews

    No.SSI-008